SUBJECT |
What is the corresponding value of RBC and WBC vs. microscopic examination? |
DATE |
2010.04.28 |
HIT |
10139 |
FILE |
|
According to CLSI guideline, following relation would be resulted on the assumption of that chemical, physical false negative or false positive results do not present.
(Count/High-Power-Field) ×4 = Count/§¡
1) RBC
Count/
High-Power-Field
|
Count/§¡
|
URiSCAN Strip Result
Plus system (semi-quantitative result)
|
Neg
|
Neg
|
Neg
|
1
|
4
|
+- (5 RBC/§¡)
|
2 ~ 3
|
8~12
|
+ (10 RBC/§¡)
|
10 ~ 13
|
40~52
|
++ (50 RBC/§¡)
|
> 60
|
> 240
|
+++ (250 RBC/§¡)
|
2) WBC
Count/
High-Power-Field
|
Count/§¡
|
URiSCAN Strip Result
Plus system (semi-quantitative result)
|
Neg
|
Neg
|
Neg
|
2 ~ 3
|
8~12
|
+- (10 RBC/§¡)
|
5 ~ 7
|
20~28
|
+ (25 RBC/§¡)
|
18 ~ 20
|
72~80
|
++ (75 RBC/§¡)
|
> 120
|
> 480
|
+++ (500 RBC/§¡)
|
|
|
|
|
|